Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (2024)

Skip Nav Destination

Article navigation

Volume 43, Issue 2

March/April 2018

  • Previous Article
  • Next Article
  • Clinical Relevance

  • SUMMARY

  • INTRODUCTION

  • METHODS AND MATERIALS

  • RESULTS

  • DISCUSSION

  • CONCLUSION

  • Acknowledgments

  • Conflict of Interest

  • REFERENCES

LABORATORY RESEARCH| March 01 2018

JK Gan;

JK Gan

Ju-Ken Gan, BDS, National University of Singapore, Faculty of Dentistry, Singapore

Search for other works by this author on:

This Site

AU Yap;

AU Yap

Adrian U. J. Yap, PhD, MSc, BDS, Grad Dip Psychotherapy, Ng Teng Fong General Hospital, Department of Dentistry, Singapore

Search for other works by this author on:

This Site

JW Cheong;

JW Cheong

Jia Wen Cheong, BDS, National University of Singapore, Faculty of Dentistry, Singapore

Search for other works by this author on:

This Site

N Arista;

N Arista

Niken Arista, BDS, National University of Singapore, Faculty of Dentistry, Singapore

Search for other works by this author on:

This Site

CBK Tan

CBK Tan

*Charles Beng Kai Tan, BDS, National University of Singapore, Faculty of Dentistry, Singapore

Search for other works by this author on:

This Site

Oper Dent (2018) 43 (2): 136–143.

  • Split-Screen
  • Views Icon Views
    • Article contents
    • Figures & tables
    • Video
    • Audio
    • Supplementary Data
    • Peer Review
  • Open the PDF for in another window
  • Tools Icon Tools
  • Search Site

Citation

JK Gan, AU Yap, JW Cheong, N Arista, CBK Tan; Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes. Oper Dent 1 March 2018; 43 (2): 136–143. doi: https://doi.org/10.2341/16-304-L

Download citation file:

  • Ris (Zotero)
  • Reference Manager
  • EasyBib
  • Bookends
  • Mendeley
  • Papers
  • EndNote
  • RefWorks
  • BibTex
toolbar search

Search Dropdown Menu

SUMMARY

This study compared the effectiveness of cure of bulk-fill composites using polywave light-emitting diode (LED; with various curing modes), monowave LED, and conventional halogen curing lights. The bulk-fill composites evaluated were Tetric N-Ceram bulk-fill (TNC), which contained a novel germanium photoinitiator (Ivocerin), and Smart Dentin Replacement (SDR). The composites were placed into black polyvinyl molds with cylindrical recesses of 4-mm height and 3-mm diameter and photopolymerized as follows: Bluephase N Polywave High (NH), 1200 mW/cm2 (10 seconds); Bluephase N Polywave Low (NL), 650 mW/cm2 (18.5 seconds); Bluephase N Polywave soft-start (NS), 0-650 mW/cm2 (5 seconds) → 1200 mW/cm2 (10 seconds); Bluephase N Monowave (NM), 800 mW/cm2 (15 seconds); QHL75 (QH), 550 mW/cm2 (21.8 seconds). Total energy output was fixed at 12,000 mJ/cm2 for all lights/modes, with the exception of NS. The cured specimens were stored in a light-proof container at 37°C for 24 hours, and hardness (Knoop Hardness Number) of the top and bottom surfaces of the specimens was determined using a Knoop microhardness tester (n=6). Hardness data and bottom-to-top hardness ratios were subjected to statistical analysis using one-way analysis of variance/Scheffe's post hoc test at a significance level of 0.05. Hardness ratios ranged from 38.43% ± 5.19% to 49.25% ± 6.38% for TNC and 50.67% ± 1.54% to 67.62% ± 6.96% for SDR. For both bulk-fill composites, the highest hardness ratios were obtained with NM and lowest hardness ratios with NL. While no significant difference in hardness ratios was observed between curing lights/modes for TNC, the hardness ratio obtained with NM was significantly higher than the hardness ratio obtained for NL for SDR.

INTRODUCTION

Bulk-fill composites are not a new notion, and many products have come and gone over the past three decades. Potential advantages of bulk-fill composites include reduction of voids in material mass as well as faster and easier placement or curing, leading to improved chairside efficiency. Polymerization shrinkage may, however, be more pronounced, and cure in deep preparations might be inadequate.1 Contemporary bulk-fill composites differ considerably from their predecessors. Through the use of novel photoinitiators, proprietary resins, special modulators, unique fillers, and filler distribution, current bulk-fill composites are claimed to have lower polymerization shrinkage and depth of cure of 4 mm or more. Studies pertaining to the depth of cure of bulk-fill composites have, however, been equivocal. While some authors have reported adequate cure at 4-mm depths,2-5 others have reported otherwise.6-8 The apparent discrepancies can be attributed to differences in testing methodologies, viscosity and translucency of the bulk-fill composites evaluated, and light-curing conditions.9

Light-cured composites set via radical photopolymerization. Light photons are absorbed by photoinitiators, and free radicals are formed in the presence of activators. The free radicals subsequently trigger the polymerization reaction, resulting in the conversion of monomers into polymers.10 Camphorquinone (CQ) is the most widely used photoinitiator and has a sensitivity peak near 470 nm in the blue range of the visible light spectrum. Because of its intense yellow color, alternative lighter-colored initiators that completely bleach out after photopolymerization have been recently promoted. These include phenyl propanedione (PPD), acyl phosphine oxide (APO), and Ivocerin. While the absorption spectrum of PPD extends from the ultraviolet (UV) wavelength range to approximately 490 nm, APO such as Lucirin TPO mainly absorbs light in the UV range. The sensitivity peak of APO is approximately 370 nm, which is considerably lower than that of CQ. Ivocerin is a newly developed germanium photoinitiator that absorbs light at a higher wavelength range than APO and has a sensitivity peak of about 420 nm.

Until the 1990s, halogen or quartz tungsten halogen curing lights were the standard curing option. Because of their wide spectral output, halogen lights required band-pass filters to limit wavelengths of light between 370 nm and 550 nm for the absorption of CQ. Their wide range of wavelengths also allows for the curing of composites employing PPD and APO as photoinitiators. However, the curing efficacy of halogen lights is low, and the high temperatures generated require external cooling and limit the lifetime of bulbs, reflectors, and filters.11 As such, the use of light-emitting diode (LED) curing lights, which emit blue light, began in the 2000s to address the problems associated with halogen curing lights. Instead of hot filaments, LED curing lights use a combination of two different doped semiconductors for light production. They consume less energy, do not require cooling fans, and have extended lifetimes without significant loss of light intensity.11 The first-generation LED curing lights had low light intensities of approximately 400 mW/cm2, while the second generation lights were able to achieve intensities of up to 1000 mW/cm2. Both first- and second-generation LED lights used only one type of LED (monowave [single-peak] technology) and were unable to cure composites with PPD and APO initiator systems. Current third-generation LED curing lights feature higher light intensities, multiple curing modes (high, low, and soft start), and avoid wavelength-compatibility issues by deploying polywave (dual/multipeak) technology.

Research pertaining to the use of polywave LED curing lights on the effectiveness of cure of bulk-fill composites is still limited.3,12,13 The curing efficacy in these studies was assessed using different methodologies, including micro-Raman spectroscopy, microhardness testing, and the ISO 4049 scraping test. Comparison of polywave and monowave LED curing lights was addressed in only one study.12 Using the ISO scraping test, no significant difference in depth of cure was found between the two LED technologies for the bulk-fill composites evaluated (Tetric Evoceram and Filtek Bulk Fill). More studies are therefore required to address the current gaps in knowledge, especially with the increased use of PPD and APO as photoinitiators, particularly in bleached shade composites. Therefore, the objective of this study was to compare the effectiveness of cure of two bulk-fill composites with polywave LED, monowave LED, and conventional halogen curing lights using microhardness testing. Curing efficacy of the high-intensity, low-intensity, and soft-start modes of the polywave LED was also appraised. It was hypothesized that no difference in the effectiveness of cure existed between the polywave LED (and its various curing modes), monowave LED, and halogen curing lights if the total light energy was kept constant.

METHODS AND MATERIALS

The technical profiles and composition of the bulk-fill composites evaluated are shown in Table 1. Tetric N-Ceram Bulk Fill (TNC; Ivoclar Vivadent, Schaan, Liechtenstein) uses CQ and Ivocerin as photoinitiators, giving it a wide absorption range of 370 nm to 460 nm.14 According to the manufacturer's instructions, TNC can be cured reliably in 4-mm increments by LED lights of 1000 mW/cm2 over 10 seconds. While TNC is a “sculptable” bulk-fill restorative, Smart Dentin Replacement (SDR; Dentsply-Caulk, Milford, DE, USA) is a “flowable” bulk-fill base material. The photoinitiator in SDR is CQ, and 4-mm increments of SDR can be cured by halogen lights of 550 mW/cm2 over 20 seconds.

Table 1

Technical Profiles of Bulk-Fill Composites Evaluated

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (2)

View large

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (3)

View Large

The composites were placed in a single increment into black polyvinyl molds with cylindrical recesses of 4-mm height and 3-mm diameter. Excess material was removed by compressing the molds between two glass slides (1-mm thick). The composites were then irradiated through the top glass slide using either a polywave LED with different curing modes (Bluephase N Polywave, Ivoclar Vivadent, Schaan, Liechtenstein), a monowave LED (Bluephase N Monowave, Ivoclar Vivadent), or a halogen (QHL75, Dentsply-Caulk) curing light. The polywave LED light offered several different curing modes including high-power, low-power, and soft-start polymerization, whereas the monowave LED light presented with only high-power curing. The exit windows of the light-curing tips were 8 mm in diameter for the three curing lights, and light intensity was verified with a radiometer (Demetron LED radiometer; Kerr Corporation, Middleton, WI, USA) prior to use to ensure consistency of energy output. The five curing light/mode combinations are detailed in Table 2. These were Bluephase N Polywave high power (NH), Bluephase N Polywave low power (NL), Bluephase N polywave soft start (NS), Bluephase N Monowave (NM), and QHL75 (QH). The total energy output (intensity × time) was standardized at 12,000 mJ/cm2 for all curing lights/modes, with the exception of NS. Because of the preset soft-start curing profile, the closest total energy achievable for NS was 13,625 mJ/cm2. Six specimens were fabricated for each composite for the various curing light/mode combinations. Immediately after light polymerization, the composite specimens were removed from their molds and stored in a light-proof container at 37°C in a humidified atmosphere for 24 hours. They were then subjected to microhardness testing with a Knoop hardness tester (FM-7, Future-Tech, Tokyo, Japan). A 10 g load was applied with a dwell time of 15 seconds to obtain the Knoop hardness number (KHN) for both top and bottom surfaces of each specimen. The KHN corresponding to each indentation was determined by measuring the dimensions of the indentations using the following formula:

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (4)

where F is the test load in kilograms and d is the longer diagonal length of an indentation in millimeters. For each surface, three readings were taken, and the mean KHN value was calculated. The KHN of the bottom was divided by the KHN of the top surface to establish the hardness ratios, which were subsequently converted to a percentage. KHN data and hardness ratios were subjected to statistical analysis using one-way Analysis of Variance and Scheffe's post hoc test at a significance level of 0.05.

Table 2

Technical Profile of Curing Lights and Modes Evaluated

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (5)

View large

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (6)

View Large

RESULTS

The mean KHN and hardness ratios (%) of the various curing light/mode combinations for TNC are shown in Table 3. The mean KHN of the top surface ranged from 22.82 ± 1.91 to 30.31 ± 4.13 when irradiated using NM and QH, respectively. The mean KHN of the bottom surface ranged from 10.32 ± 1.04 to 13.43 ± 1.52 for NH and QH, respectively. The hardness ratio for TNC ranged from 38.43% ± 5.19% to 49.25% ± 6.38% for NL and NM, respectively. Results of statistical analysis for TNC are reflected in Table 4. At the top surface, the KHNs of QH and NL were significantly higher than NM, while for the bottom surface, the KHN of QH was significantly higher than that of NH. However, no significant difference in hardness ratio was observed.

Table 3

Mean Top KHN, Bottom KHN, and Hardness Ratio (%) With the Different Lights/Modes for TNC and SDR

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (7)

View large

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (8)

View Large

Table 4

Statistical Comparison of Top KHN, Bottom KHN, and Hardness Ratio (%) for TNC and SDR

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (9)

View large

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (10)

View Large

The mean KHN and hardness ratios (%) of the various curing light/mode combinations for SDR are also shown in Table 3. The mean top KHN ranged from 15.21 ± 2.42 to 18.35 ± 1.41 for NH and QH, while the mean bottom KHN ranged from 8.30 ± 0.69 to 10.95 ± 1.43 for NL and NM, respectively. The hardness ratio for SDR ranged from 50.67% ± 1.54% to 67.62% ± 6.96% for NL and NM, respectively. Results of the statistical analysis for SDR are reflected in Table 4. At the top surface, the KHN of QH was significantly higher than that of NH. At the bottom surface, the KHNs for NM and QH were significantly higher than that of NL. The bottom KHN of NM was also significantly higher than that of NH. Unlike TNC, significant differences in hardness ratio for SDR were observed. The hardness ratio obtained with NM was significantly higher than with NL.

DISCUSSION

The effectiveness of cure of bulk-fill composites with polywave LED, monowave LED, and conventional halogen curing lights was evaluated. The null hypothesis was rejected as significant differences in effectiveness of cure existed between the different lights and curing modes despite regulating the total light energy. Curing efficacy can be assessed by direct and indirect methods. Direct methods, such as infrared and Raman spectroscopy, are not routinely used as they are complex, expensive, and time-consuming to perform.15 Indirect methods include visual appraisal, scraping (ISO 4049), and hardness testing. While visual appraisal does not offer scientific objectivity, the ISO scraping test generally results in greater depths of cure when compared with hardness testing.6 Hardness is an indicator of the degree of polymerization, and a good correlation between Knoop hardness and infrared spectroscopy has been reported.16,17 In view of its relative efficiency and popularity, Knoop hardness testing was selected to determine the effectiveness of composite cure. Hardness testing was done 24 hours after photopolymerization to allow for composite postcure.18

Adequate photopolymerization is essential for optimization of physicomechanical properties and clinical longevity of composite restorations.19 In addition, inadequately cured composites are also cytotoxic because of residual monomers and other reactive components.20,21 Composite restorations should ideally be equally cured throughout. The bottom-to-top hardness ratio of the 4-mm-thick bulk-fill specimens should approximate or equal 1 (100%). Many studies have, however, used a hardness ratio of 0.8 or 80% as the standard for satisfactory cure due to material and light-curing constraints.22,23 Material factors affecting photopolymerization involve thickness, shade, opacity, and composition, while those related to curing lights include light intensity, wavelength, exposure time, size, location, and orientation of the light probes.19 As thickness, materials, total light energy (intensity Ă— time), and light probe variables including curing distance were controlled during the experiment, the results can be largely attributed to light type, spectral wavelengths, and light intensity modification during curing (ie, continuous versus soft-start curing).

At both top and bottom surfaces, the KHN of TNC was higher than SDR regardless of curing light/mode. Results corroborated those of similar studies concerning these materials and can be attributed to TNC's higher filler content when compared with SDR.6,7 While TNC can be placed up to the surface of the overall composite resin restoration and functionally loaded, SDR requires a “capping layer” of conventional composite to sustain functional loads because of its lower filler loading. At the top surface, significant differences in KHN were observed between curing lights/modes. For TNC, curing with the halogen light and polywave LED (NL in particular) resulted in harder top surfaces than with the monowave LED light (NM). This finding may be attributed to the narrower spectral output of the NM. For SDR, significant differences in the top KHN was observed between the halogen (QH) and polywave LED in high-power mode (NH). The harder surface associated with QH could be contributed in part to a thermal effect and longer exposure time as emission spectrum and total energy output were similar.24 Exposure time for QH was approximately double that of NH (21.8 vs 10 seconds), and heating of composites from halogen lights has been shown to increase hardness.25

As light passes through the bulk-fill composites, intensity is clearly reduced because of light absorption and scattering by the materials, attenuating the potential for photopolymerization.26 At the bottom surface, photopolymerization of TNC with the halogen light (QH) resulted in significantly higher bottom KHN than with the polywave LED at high power (NH). Thermal effects are negligible at the bottom surface of restorations as composites are poor conductors of heat.27 As emission spectrum and total energy output were comparable, differences may well be due to variations in light attenuation, light exposure time, and the ensuing polymeric network type. Light attenuation with NH may be greater, especially considering its short curing time. Previous studies have reported the need for longer curing times with LED lights when compared with halogen lights for achieving a similar depth of cure and mechanical properties.28,29 Hardness of composites is dependent not only on the degree of conversion but also on the nature of and bonding between monomers.16 Polymers with more cross-linked chains are harder than those with linear chains.30 Theoretically, the use of high light intensity in the initial phase of curing should result in a greater number of growth centers and higher cross-link density.31 The aforementioned may, however, be mitigated by the short curing time associated with NH. As such, the cross-link density of bulk-fill composites associated with the various lights/modes requires further investigation.

For SDR, the bottom KHN with the halogen light (QH) was significantly higher than the polywave LED in low-power mode (NL). Significant differences in bottom hardness were also observed between the LED lights. Photopolymerization with the monowave light (NM) resulted in significantly higher bottom KHN than curing with the polywave LED in both high- and low-power modes (NH and NL). The incongruence in findings when compared with TNC may be attributed to variances in composite composition. In addition to photoinitiator and resin differences, SDR has a lower filler loading and larger filler particle sizes than TNC, resulting in a more translucent material. Monomer conversion has been found to be inversely proportional to filler loading owing to decreased light transmission.32 Light scattering from smaller filler particles has been found to reduce depth of cure, especially when filler sizes are similar to the wavelength of the emitted light.33 Despite comparable curing times, photopolymerization with the halogen light still resulted in harder bottom surfaces than with the polywave LED light, reinforcing the necessity for cross-link density studies. Photopolymerization with the monowave LED light resulted in significantly higher bottom surface KHN than the polywave light, with the exception of the soft-start mode. The better performance of the polywave soft-start mode could be attributed to its slightly higher total energy (13,625 mJ/cm2). The total energy of the soft-start mode could not be harmonized to 12,000 mJ/cm2 because of the manufacturer's programmed settings. The polywave LED curing offered no advantage over its monowave counterpart as SDR uses primarily CQ as its photoinitiator.

For both TNC and SDR, the hardness ratio was lower than 80% for all curing lights/modes. This can be attributed to attenuated irradiance reaching the bottom surfaces.26 The highest hardness ratio achieved was 49.25% ± 6.38% and 67.62% ± 6.96% for TNC and SDR, respectively. Results corroborated a recent independent study by Yap and others involving the same materials.6 To achieve a hardness ratio of 80%, TNC and SDR need to be limited to increments of 2.5 mm and 3 mm correspondingly.6 Other authors have, however, reported a bottom-to-top hardness ratio of 80% or more.34,35 The divergence in outcomes can be attributed to differences in bulk-fill materials evaluated, curing light, or parameters and methodologies employed. For both bulk-fill composites, the highest hardness ratio was obtained with NM and lowest with NL despite their identical total energy output. Ranking of hardness ratios differed slightly between TNC and SDR and were as follows: TNC – NM > QH > NS > NH > NL; SDR – NM > NS > NH > QH > NL. When only the LED curing lights were considered, the ranking of hardness ratio was similar for both bulk-fill composites. Photopolymerization with monowave LED gave the highest hardness ratio followed by the polywave LED light in soft-start, high-power, and low-power modes. The monowave LED light thus appears to be somewhat more effective than the polywave LED at 4-mm depths.

For TNC, no significant difference in hardness ratio was observed between the various curing lights/modes. Results supported those of Meenes and others12 using the ISO 4049 scraping test and customized tooth molds. They found the influence of curing lights to be insignificant, but there was a relatively significant interaction between materials and mold types on composite depth of cure. The use of stainless steel molds led to a deeper depth of cure for Tetric Evoceram bulk-fill, rationalizing the use of unreflective black polyvinyl molds in the present study. For SDR, photopolymerization with the monowave LED resulted in a significantly higher hardness ratio than with the polywave LED at low-power mode despite standardization of the total energy (67.62% vs 50.67%). The irradiance of the polywave LED light even with a lower power mode was already 650 mW/cm2, well above the traditionally recommended minimum light intensity of 400 mW/cm2.36 The 17% variance in hardness ratio between the two LED lights is of concern and warrants further investigation. It may be attributed in part to the lower percentage of light transmitted through composites offered by polywave LED lights when compared with monowave LED lights.37 Clinically, the reduction in hardness ratio may be even higher as the curing light probe may be 8 mm or more from the composite surface. The latter has been shown to significantly reduce the degree of conversion at the bottom surface of restorations.38 The effect of light probe distance on curing efficacy of poly- and monowave LED curing lights should also be explored.

CONCLUSION

Within the limitations of this in vitro study, the following conclusions can be made:

  • Top and bottom surface hardness achieved with the halogen curing light was usually higher than that obtained with the LED lights when total energy was controlled.

  • The ideal bottom-to-top hardness ratio of 0.8 (80%) was not achieved by either bulk-fill composite evaluated regardless of curing lights/modes.

  • For both bulk-fill composites, the highest hardness ratios were obtained with the monowave LED curing light and lowest with the polywave LED curing light in low-power mode.

  • Photopolymerization with monowave LED lights may be more effective than with polywave LED lights for camphorquinone-based bulk-fill composites.

Acknowledgments

This work was supported by the UROP grant from the Faculty of Dentistry, National University of Singapore. The authors would like to thank the manufacturers for their materials support and loan of curing units.

Conflict of Interest

The authors have no proprietary, financial, or other personal interest of any nature or kind in any product, service, and/or company that is presented in this article.

REFERENCES

1

Christensen

GJ

(

2012

)

Advantages and challenges of bulk-fill resins

Clinicians Report

5

(

1

)

1

-

2

.

2

Zorzin

J,

Maier

E,

Harre

S,

Fey

T,

Belli

R,

Lohbauer

U,

Petschelt

A,

&

Taschner

M

(

2015

)

Bulk-fill resin composites: polymerization properties and extended light curing

Dental Materials

31

(

3

)

293

-

301

.

3

Li

X,

Pongprueksa

P,

Van Meerbeek

B,

&

De Munck

J

(

2015

)

Curing profile of bulk-fill resin-based composites

Journal of Dentistry

43

(

6

)

664

-

72

.

4

Goracci

C,

Cadenaro

M,

Fontanive

L,

Giangrosso

G,

Juloski

J,

Vichi

A,

&

Ferrari

M

(

2014

)

Polymerization efficiency and flexural strength of low-stress restorative composites

Dental Materials

30

(

6

)

688

-

94

.

5

Bucuta

S,

&

Ilie

N

(

2014

)

Light transmittance and micro-mechanical properties of bulk fill vs. conventional resin based composites

Clinical Oral Investigations

18

(

8

)

1991

-

2000

.

6

Yap

AU,

Pandya

M,

&

Toh

WS

(

2016

)

Depth of cure of contemporary bulk-fill resin-based composites

Dental Materials Journal

35

(

3

)

503

-

10

.

7

Jang

JH,

Park

SH,

&

Hwang

IN

(

2015

)

Polymerization shrinkage and depth of cure of bulk-fill resin composites and highly filled flowable resin

Operative Dentistry

40

(

2

)

172

-

80

.

8

Garcia

D,

Yaman

P,

Dennison

J,

&

Neiva

G

(

2014

)

Polymerization shrinkage and depth of cure of bulk fill flowable composite resins

Operative Dentistry

39

(

4

)

441

-

8

.

9

Ilie

N,

&

Stark

K

(

2015

)

Effect of different curing protocols on the mechanical properties of low-viscosity bulk-fill composites

Clinical Oral Investigations

19

(

2

)

271

-

9

.

10

Stansbury

JW

(

2000

)

Curing dental resins and composites by photopolymerization

Journal of Esthetic Dentistry

12

(

6

)

300

-

8

.

11

Krämer

N,

Lohbauer

U,

GarcĂ­a-Godoy

F,

&

Frankenberger

R

(

2008

)

Light curing of resin-based composites in the LED era

American Journal of Dentistry

21

(

3

)

135

-

42

.

12

Menees

TS,

Lin

CP,

Kojic

DD,

Burgess

JO,

&

Lawson

NC

(

2015

)

Depth of cure of bulk fill composites with monowave and polywave curing lights

American Journal of Dentistry

28

(

6

)

357

-

61

.

13

AlShaafi

MM,

Haenel

T,

Sullivan

B,

Labrie

D,

Alqahtani

MQ,

&

Price

RB

(

2016

).

Effect of a broad-spectrum LED curing light on the Knoop microhardness of four posterior resin based composites at 2, 4 and 6-mm depths

Journal of Dentistry

45

14

-

8

.

14

Moszner

N,

Fischer

U,

Ganster

B,

Liska

R,

&

Rheinberger

V

(

2008

)

Benzoyl germanium derivatives as novel visible light photoinitiators for dental materials

Dental Materials

24

(

7

)

901

-

7

.

15

Rueggeberg

FA,

&

Craig

RG

(

1988

)

Correlation of parameters used to estimate monomer conversion in a light-cured composite

Journal of Dental Research

67

(

6

)

932

-

7

.

16

Asmussen

E

(

1982

)

Factors affecting the quantity of remaining double bonds in restorative resin polymers

Scandinavian Journal of Dental Research

90

(

6

)

490

-

6

.

17

Dewald

JP,

&

Ferracane

JL

(

1987

)

A comparison of four modes of evaluating depth of cure of light-activated composites

Journal of Dental Research

66

(

3

)

727

-

30

.

18

Al-Ahdal

K,

Ilie

N,

Silikas

N,

&

Watts

DC

(

2015

)

Polymerization kinetics and impact of post polymerization on the degree of conversion of bulk-fill resin-composite at clinically relevant depth

Dental Materials

31

(

10

)

1207

-

13

.

19

Jadhav

S,

Hegde

V,

Aher

G,

&

Fajandar

N

(

2011

)

Influence of light curing units on failure of direct composite restorations

Journal of Conservative Dentistry

14

(

3

)

225

-

7

.

20

Toh

WS,

Yap

AU,

&

Lim

SY

(

2015

)

In vitro biocompatibility of contemporary bulk-fill composites

Operative Dentistry

40

(

6

)

644

-

52

.

21

Caughman

WF,

Caughman

GB,

Shiflett

RA,

Rueggeberg

F,

&

Schuster

GS

(

1991

)

Correlation of cytotoxicity, filler loading and curing time of dental composites

Biomaterials

12

(

8

)

737

-

40

.

22

Watts

DC,

Amer

O,

&

Combe

EC

(

1984

)

Characteristics of visible light-cured composite systems

British Dental Journal

156

(

6

)

209

-

15

.

23

Yap

AUJ,

Soh

MS,

&

Siow

KS

(

2002

)

Effectiveness of cure with pulse activation and soft-start polymerization

Operative Dentistry

27

(

1

)

44

-

9

.

24

Yap

AU,

&

Soh

MS

(

2003

)

Thermal emission by different light-curing units

Operative Dentistry

28

(

3

)

260

-

6

.

25

Torno

V,

Soares

P,

Martin

JM,

Mazur

RF,

Souza

EM,

&

Vieira

S

(

2008

)

Effects of irradiance, wavelength, and thermal emission of different light curing units on the Knoop and Vickers hardness of a composite resin

Journal of Biomedical Materials Research: Part B, Applied Biomaterials

85

(

1

)

166

-

71

.

26

Ruyter

IE,

&

Oysaed

H

(

1982

)

Conversion in different depths of ultraviolet and visible light activated composite materials

Acta Odontologica Scandinavica

40

(

3

)

179

-

92

.

27

Civjan

S,

Barone

JJ,

Reinke

PE,

&

Selting

WJ

(

1972

)

Thermal properities of non-metallic restorative materials

Journal of Dental Research

51

(

4

)

1030

-

7

.

28

Besnault

C,

Pradelle-Plasse

N,

Picard

B,

&

Colon

P

(

2003

)

Effect of a LED versus halogen light cure polymerization on the curing characteristics of three composite resins

American Journal of Dentistry

16

(

5

)

323

-

8

.

29

Nomoto

R,

McCabe

JF,

&

Hirano

S

(

2004

)

Comparison of halogen, plasma and LED curing units

Operative Dentistry

29

(

3

)

287

-

94

.

30

Borges

BCD,

Groninger

AIS,

Soares

GP,

Santos-Daroz

CBD,

Ambrosano

GMB,

Marchi

GM,

Giorgi

MCC,

&

Aguiar

FHB

(

2012

)

Curing quality of composites as influenced by the filler content, light source and curing time

Dentistry

1

103

.

31

Filho

JDN,

Poskus

LT,

GuimarĂŁes

JGA,

Barcellos

AAL,

&

Silva

EM

(

2008

)

Degree of conversion and plasticization of dimethacrylate-based polymeric matrices: influence of light-curing mode

Journal of Oral Science

50

(

3

)

315

-

321

.

32

Howard

B,

Wilson

ND,

Newman

SM,

Pfeifer

CS,

&

Stansbury

JW

(

2010

)

Relationships between conversion, temperature and optical properties during composite photopolymerization

Acta Biomaterialia

6

(

6

)

2053

-

2059

.

33

Bouschlicher

MR,

Rueggeberg

FA,

&

Wilson

BM

(

2004

)

Correlation of bottom-to-top surface microhardness and conversion ratios for a variety of resin composite compositions

Operative Dentistry

29

(

6

)

698

-

704

.

34

Kim

EH,

Jung

KH,

Son

SA,

Hur

B,

Kwon

YH,

&

Park

JK

(

2015

)

Effect of resin thickness on the microhardness and optical properties of bulk-fill resin composites

Restorative Dentistry & Endodontics

40

(

2

)

128

-

35

.

35

Alshali

RZ,

Salim

NA,

Satterthwaite

JD,

&

Silikas

N

(

2015

)

Post-irradiation hardness development, chemical softening, and thermal stability of bulk-fill and conventional resin-composites

Journal of Dentistry

43

(

2

)

209

-

18

.

36

Rueggeberg

FA,

Caughman

WF,

Curtis

JW

Jr
(

1994

)

Effect of light intensity and exposure duration on cure of resin composite

Operative Dentistry

19

(

1

)

26

-

32

.

37

Kolpakova

&

Anna

V

(

2015

)

Comparison of monowave and polywave transmission and curing profile thru various composites

University Honors Thesis, Portland State University

.

38

Catelan

A,

Mainardi Mdo

C,

Soares

GP,

de Lima

AF,

Ambrosano

GM,

Lima

DA,

Marchi

GM,

&

Aguiar

FH

(

2014

)

Effect of light curing protocol on degree of conversion of composites

Acta Odontologica Scandinavica

72

(

8

)

898

-

902

.

Operative Dentistry, 2018

2018

1,338 Views

20 Web of Science

27 Crossref

View Metrics

Citing articles via

Google Scholar

CrossRef (27)

Get Email Alerts

Article Activity Alert

Latest Issue Alert

Close Modal

Bulk-Fill Composites: Effectiveness of Cure With Poly- and Monowave Curing Lights and Modes (2024)

FAQs

What is the effect of shortened light curing modes on bulk fill resin composites? ›

Shortened light curing does not affect volumetric polymerization shrinkage or cohesive tensile strength but negatively affects the shear bond strength of some bulk-fill resin composites.

Do dual cure bulk fill resin composites reduce gaps and improve depth of cure? ›

Dual-cure bulk-fill composites did not reduce gaps compared to light-cure bulk-fill, but they can improve depth of cure of bulk-filled restorations.

How long does it take to cure composite resin with light? ›

For all the materials, the hardness number of composite resin after 20- and 40-sec exposure to light were always significantly lower than the 60-sec exposure.

What curing lamp is used for composites? ›

A dental curing light is a piece of dental equipment that is used for polymerization of light-cure resin-based composites. It can be used on several different dental materials that are curable by light. The light used falls under the visible blue light spectrum.

What are the disadvantages of bulk-fill composite? ›

Polymerization shrinkage is the major disadvantage of using composites (2,5,6), since it creates stresses between tooth and restoration, leading to failures at the adhesive interface, microgaps and cuspid deflection (7,8,9).

What are the advantages of bulk-fill composites? ›

Because bulk fills can be completed in fewer layers than conventional composite, they can actually make the placement process faster and easier. This, in turn, shortens chair time, which is particularly appreciated by clinicians treating restless or pediatric patients.

What is the depth of cure of bulk-fill composite? ›

Bulk-fill composite resins, which can be cured at a thickness of 4-5 mm, were introduced to reduce the duration of restorative procedures [ 4 ]. Bulk-fill composite resins are cured in thicknesses greater than 2 mm due to some mechanisms.

What are the types of bulk-fill resin composites? ›

Bulk-fill RBC restorative materials can be categorized into high-viscosity or low- viscosity, light or dual cured. Table 1 discusses some of these available materials, their maximum incremental depth and whether or not they require a conventional RBC capping layer.

How do bulk-fill composite resins differ from conventional composite resins? ›

Base bulk-fill composites have low viscosity, allowing for their placement and adaptation in deep cavities. However, their lower filler content, which results in lower wear resistance, requires the base of the bulk-fill to be covered with a conventional composite (two-step bulk technique).

What is the ideal curing time for composite? ›

The same researchers reported that light curing of a preheated composite for 5 sec resulted in a greater degree of polymerization than light curing at room temperature for 40 sec.

What happens if you cure UV resin too long? ›

If you cure UV resin for too long, it can become brittle and yellow over time. Over-curing can also cause the resin to crack or shrink.

Can you speed up resin cure time? ›

Place a space heater or heat lamp near your piece to encourage the resin to cure faster. đź’ˇTIP: Be aware to keep temperature constant for the first 24 hours of curing. If there is a temperature drop, you can end up with things like dimples or an orange peel effect in the cured resin.

What is the effect of light curing protocols on the mechanical behavior of bulk-fill resin composites? ›

The higher irradiance in a shorter time interval compromises the mechanical behavior of the resin composites, which may result in undesirable clinical outcomes.

How does intensity of the curing light affect depth of cure? ›

Increments of light-curing composite resin should not exceed 2 mm; optimal hardness only developed after one hour; at a depth of 2 mm an increase in light intensity produced a significant increase in Vickers Hardness; beyond 2 mm depth the increased light intensity did not produce a significant increase in Vickers ...

What effect of preheating on the cytotoxicity of bulk-fill composite resins? ›

Pre-heating of bulk-fill composite resin did not affect their cytotoxicity.

How does pre heating affect the microhardness and depth of cure of bulk-fill composite resins? ›

The preheating procedure increased the microhardness values of all bulk-fill composites. The composition of the tested composite resins affected their microhardness values.

Top Articles
Latest Posts
Article information

Author: Prof. Nancy Dach

Last Updated:

Views: 5429

Rating: 4.7 / 5 (77 voted)

Reviews: 84% of readers found this page helpful

Author information

Name: Prof. Nancy Dach

Birthday: 1993-08-23

Address: 569 Waelchi Ports, South Blainebury, LA 11589

Phone: +9958996486049

Job: Sales Manager

Hobby: Web surfing, Scuba diving, Mountaineering, Writing, Sailing, Dance, Blacksmithing

Introduction: My name is Prof. Nancy Dach, I am a lively, joyous, courageous, lovely, tender, charming, open person who loves writing and wants to share my knowledge and understanding with you.